All research

Biomedical devices & simulation · 2024

ExoventQ: a low-cost portable negative-pressure ventilator

A student-led respiratory support prototype with continuous and cyclic negative-pressure modes, closed-loop pressure control and safety alarms.

BSc Electrical Engineering senior design project, 2020–2021 · Qatar University QUST-1-CENG-2023-796 · Prince Sattam Bin Abdulaziz University project PSAU/2024/R/1445

Illustration of the system. A patient lies in a pressure vessel connected through a three-way valve to a vacuum pump. A pressure sensor feeds a control circuit linked to a touchscreen interface.
Figure 1 from Gaben et al. (2024). The negative-pressure vessel, three-way valve and pump, with pressure feedback to the control circuit and user interface.View full-size figureSource articleCC BY 4.0

Approach

Can a portable negative-pressure ventilator be built from affordable, readily available components?

The COVID-19 pandemic exposed global shortages of mechanical ventilators. ExoventQ began as a final-year engineering project to explore a lower-cost alternative that applies negative pressure around the chest instead of delivering positive pressure through the airway. The team designed the pressure vessel, pumping system, control electronics and clinician interface, then tested the integrated prototype.

  1. Design the pressure vessel
  2. Build pumping and control
  3. Test modes and safety

Designing and building the pressure vessel

IEEE Access · 2024

The team first built a wooden prismatic vessel to test operation and sealing, then developed a transparent polycarbonate vessel for the final prototype. ANSYS stress analysis compared semi-cylindrical and rectangular designs and identified a squircle-faced semi-cylinder as the most suitable shape for limiting structural failure.

Polycarbonate was chosen over acrylic because acrylic is more likely to crack during drilling or impact. Neoprene neck and torso seals were selected over latex for their flexibility and lower allergy risk, and the squircle shape avoids tilting the suction tube and valve.

Vessel vacuum design limit
50 mbar
Polycarbonate wall thickness
3 mm
Adult male design range
75–90 kg
Vessel prototypes described in Section III-C of Gaben et al. (2024).
Wooden test vesselPolycarbonate vessel
Length0.90 m0.90 m
Width0.64 m0.76 m
Height0.415 m0.415 m
Main wall12 mm plywood3 mm polycarbonate
SupportBattens and 25 mm girderGalvanised steel frame
A 3D model of the rounded, squircle-faced vessel above a photograph of the assembled transparent polycarbonate vessel with its supporting frame.
Figure 6 from Gaben et al. (2024). 3D model (a) and implemented polycarbonate vessel (b).View full-size figureSource articleCC BY 4.0

Generating and controlling negative pressure

IEEE Access · 2024

A 200 W vacuum-cleaner motor-blower provided suction, with a maximum vacuum of 15 kPa and airflow of about 13 L/s. A 3D-printed three-way butterfly valve, moved by a servo motor, regulated the air drawn from the vessel.

An Arduino Uno read a pressure sensor checked against a 2 m water manometer and adjusted the valve whenever vessel pressure moved more than 0.5 cmH₂O from its target. A DHT22 sensor monitored temperature, and a Raspberry Pi touchscreen provided the settings, readings and live pressure waveform.

Pressure correction threshold
±0.5 cmH₂O
Maximum pump vacuum
15 kPa
Pump airflow
~13 L/s
Clinician-adjustable settings described in Section III of Gaben et al. (2024).
Range
Inspiration pressure15–35 cmH₂O
Expiration pressure0–15 cmH₂O
Respiratory rate10–30 breaths per minute
Inspiration-to-expiration ratio1:1 to 1:4
Continuous negative pressure5–15 mbar vacuum
Control system diagram. A clinician uses the interface connected to a control circuit, which drives the motor and valve actuator while a pressure sensor feeds back the vessel pressure.
Figure 7 from Gaben et al. (2024). Layout of the control system.View full-size figureSource articleCC BY 4.0
A 3D model of the three-way butterfly valve above two photographs of the 3D-printed valve, including one connected to the vacuum hose.
Figure 8 from Gaben et al. (2024). 3D model (a) and 3D-printed three-way butterfly valve (b, c).View full-size figureSource articleCC BY 4.0

Testing ventilation modes and safety features

IEEE Access · 2024

The prototype was tested in cyclic negative-pressure ventilation and continuous negative extrathoracic pressure modes. Pressure curves were recorded while changing the target pressures, respiratory rate and inspiration-to-expiration ratio, and the full system was then checked through the touchscreen interface.

A moderate alarm turns the interface button orange when pressure deviates by about 5 units from the setting. A severe alarm, at about 10 units, turns it red and sounds a buzzer. A relay-based safety switch stops the motor if the vessel vacuum falls below 40 mbar.

Ventilation modes tested
2
Measured inspiration pressure at a 20 cmH₂O setting
19.66 cmH₂O
Safety cut-off threshold
40 mbar
Interface readings shown in Figures 12 and 13 of Gaben et al. (2024).
Set pressureDisplayed pressure
Cyclic inspiration20 cmH₂O19.66 cmH₂O
Cyclic expiration5 cmH₂O4.54 cmH₂O
Continuous mode5 cmH₂O4.91 cmH₂O
Moderate alarm test25 cmH₂O20.43 cmH₂O
Severe alarm test30 cmH₂O19.75 cmH₂O
Three pressure-time plots showing repeating inspiration and expiration waves under different pressure, respiratory rate and timing settings.
Figure 10 from Gaben et al. (2024). Cyclic pressure curves for different settings.View full-size figureSource articleCC BY 4.0
Two pressure-time plots in continuous mode. Pressure rises quickly and then stays close to 5 cmH₂O in one plot and 10 cmH₂O in the other.
Figure 11 from Gaben et al. (2024). Continuous-mode pressure curves at two expiration pressure settings.View full-size figureSource articleCC BY 4.0
Two photographs of the touchscreen interface during testing, one showing a cyclic pressure waveform and the other a nearly constant pressure trace.
Figure 12 from Gaben et al. (2024). Full-system tests in cyclic (a) and continuous (b) modes.View full-size figureSource articleCC BY 4.0
Two photographs of the touchscreen during alarm tests. The alarm button is orange in the moderate condition and red in the severe condition.
Figure 13 from Gaben et al. (2024). Moderate (a) and severe (b) alarm tests.View full-size figureSource articleCC BY 4.0

My contribution

ExoventQ was developed by a student group as their final-year BSc Electrical Engineering senior design project. I supported the project through engineering analysis, prototype evaluation and interpretation of the test results, contributed to other project tasks, and helped prepare and revise the manuscript.

Study scope

This is an engineering prototype and was not tested on people. Mannequins were used to evaluate the neck and torso seals, but their effect on vessel pressure was not examined in depth. Detailed risk analysis, long-term reliability, user training and clinical studies are needed before patient use.